Perguntas populares do fórum
Fazer uma perguntaNão há nenhuma pergunta. Seja o primeiro a fazer uma pergunta!
Documentos
Background
The BiPAP Vision Ventilatory Support System, which was developed by Phillips Respironics, is a ventilation machine mainly for patients with respiratory failures. It delivers bilevel positive airway pressure (BiPAP), providing two distinct levels of pressure: a higher pressure for inhalation and a lower pressure for exhalation. This helps improve oxygenation, reduce the work of breathing, and prevent respiratory fatigue.
The BiPAP Vision Ventilatory Support System was produced by Philips Respironics, a company known for its innovations in sleep and respiratory care. The BiPAP was first developed in the early 1980s, aimed at providing an alternative for patients with respiratory issues. Philips Respironics was the main producer in this field, enhancing the technology over the years to improve usability and effectiveness.
The BiPAP Vision, introduced in the early 2000s, incorporated more advanced features such as customizable settings, enhanced monitoring capabilities, and user-friendly interfaces. This model was designed to meet the needs of all users, facilitating better management of various respiratory conditions. With the BiPAP Vision Ventilatory Support System repair, we can fix all of the components in the system.
Identification
The BiPAP Vision Ventilatory Support System is usually labeled as "BIPAP Vision" on the front or sides of the machine. The machine is a compact boxlike shape with a display screen and buttons on it for adjustments. Some models have ports for connecting to other devices for data transferring. There is also a serial number on the bottom of the device for identification.
Technical Specifications
- Controls
- IPAP
- 4 to 40 cm H20
- EPAP
- 4 to 20 cm H20
- CPAP
- 4 to 20 cm H20
- Rate
- 4 to 40 BPM
- Timed Inspiration
- 0.5 to 3.0 sec
- IPAP Rise Time
- 0.05 to 0.40 sec
- Oxygen Concentration
- 21 to 100%
- IPAP
- Display
- IPAP
- 0 to 50cm H20
- EPAP
- 0 to 50 H20
- CPAP
- 0 to 50 H20
- Rate
- 0 to 150 bpm
- Exhaled Estimated Tidal Volume
- 0 to 4000 ml
- Minute Ventilation
- 0 to 99 L/min
- Total leak
- 0 to 300 L/min
- Patient leak
- 0 to 300 L/min
- Peak Inspiratory Pressure
- 0 to 50 cm H20
- Precent of Patient Triggered Breaths
- 0 to 100%
- Ti
- 0 to 100%
- IPAP
- Alarms
- High Pressure
- 0 to 50 cm H20
- Low Pressure
- Disabled, 1 to 40 cm H20
- Low Pressure Delay
- 0 to 60 sec
- Apnea
- Disabled 20 to 40 sec
- Low Minute Ventilation
- Disabled, 1 to 99 L/min
- High Rate
- 4 go 120 BPM
- Low Rate
- 4 to 120 BPM
- High Pressure
Source: BiPAP Vision Ventilatory Support System