How to Test the Valve System on a Fisher Paykel Neopuff
Introdução
Ir para o passo 1Please refer to page 7 of the device’s Technical Manual for source information.
-
-
Set the gas supply to 5 L/min.
-
-
-
Reset the peak inspiratory pressure to 20 cmh2O and turn off the gas flow. Testing is now complete. If the F&P Neopuff/Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers).
-
To reassemble your device, follow these instructions in reverse order.
To reassemble your device, follow these instructions in reverse order.