Philips Respironics V60 Fan Replacement

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.

O que você precisa

  1. Philips Respironics V60 Fan Replacement, shut down the ventilator: passo 1, imagem 1 %32 Philips Respironics V60 Fan Replacement, shut down the ventilator: passo 1, imagem 2 %32
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

    Pergunte ao FixBot

    Comentar

    • Disconnect the ventilator from the AC power.

    Pergunte ao FixBot

    Comentar

    • Unplug the ventilator

    Pergunte ao FixBot

    Comentar

  2. Philips Respironics V60 Fan Replacement: passo 4, imagem 1 %32
    • Disconnect the internal battery from the power harness connector.

    Pergunte ao FixBot

    Comentar

  3. Philips Respironics V60 Fan Replacement, Top Cover: passo 5, imagem 1 %32
    • Turn the captive D-ring fastener on the bottom of the panel counter-clockwise one-quarter turn and release.

    • Then remove the right side panel.

    Pergunte ao FixBot

    Comentar

  4. Philips Respironics V60 Fan Replacement: passo 6, imagem 1 %32
    • Remove the 4 screws using a screwdriver.

    • Remove the top cover.

    Pergunte ao FixBot

    Comentar

  5. Philips Respironics V60 Fan Replacement, Fan: passo 7, imagem 1 %32
    • Remove the fan mount push pins that attach the fan to the back panel.

    • Pliers may be required.

    Pergunte ao FixBot

    Comentar

  6. Philips Respironics V60 Fan Replacement: passo 8, imagem 1 %32
    • Disconnect fan cable from motor controller.

    Pergunte ao FixBot

    Comentar

  7. Philips Respironics V60 Fan Replacement: passo 9, imagem 1 %32
    • Remove fan,

    • from back panel.

    Pergunte ao FixBot

    Comentar

  8. Philips Respironics V60 Fan Replacement: passo 10, imagem 1 %32
    • Remove fan filter assembly (the fan filter and housing), from outside the back panel.

    • When installing, orient the fan with the fan cable at the top inside of the fan, and orient the fan filter and housing as shown.

    Pergunte ao FixBot

    Comentar

Conclusão

To reassemble your device, follow these instructions in reverse order.

Cancelar: não concluí este guia.

Uma outra pessoa concluiu este guia.

Rie Kuraoka

Membro desde: 01/23/20

913 Reputação

Autoria de 7 guias

Equipe

Cal Poly, Team S7-G10, Paton Spring 2020 Membro de Cal Poly, Team S7-G10, Paton Spring 2020

CPSU-PATON-S20S7G10

2 Membros

Autoria de 9 guias

0Comentários do guia

Comentar

Visualizações:

Últimas 24 horas: 2

Últimos 7 dias: 3

Últimos 30 dias: 24

Ao todo: 1,034