Ir para o conteúdo principal

Viasys Avea UIM and Plastic Cover Disassembly

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.

O que você precisa

  1. Viasys Avea UIM and Plastic Cover Disassembly, Viasys Avea UIM and Plastic Cover Disassembly: passo 1, imagem 1 %32
    • Remove the ‘rubber collar’ located at the UIM rear neck, by grasping one of the two rubber tabs at the bottom. Pull firmly in an “arcing” motion.

    • Pull UIM forward and look for the Phillips Screw on the front arm cover. Remove (1) Phillips screw from the front arm cover. Remove the front arm cover.

    • Remove the two mounting screws now visible inside the ‘back arm cover’. Tilt the UIM down and remove the ‘back arm cover’.

  2. Viasys Avea UIM and Plastic Cover Disassembly: passo 4, imagem 1 %32
    • Remove the (4) 11/32 KEP nuts that secure the ‘plastic top cover’ opening onto the chassis.

    • Remove (2) Phillips screws that secure the ‘plastic top cover’ to the rear chassis.

    • Slide the top ‘plastic top cover’ upward and forward away from the chassis.

    • Remove the (2) Phillips screws and washers that secure the HSSC Cable on the GDE. Unplug the HSSC Cable.

    • Carefully feed the HSSC Cable through the ‘plastic top cover’ opening.

    • While continuously supporting the UIM, remove (4) 3/8 nuts that fasten the UIM support arm onto the pneumatics.

    • Lift UIM off of pneumatics and set aside.

    • Remove plastic cover and set aside.

Conclusão

To reassemble your device, follow these instructions in reverse order.

Anthony Hernandez

Membro desde: 05/04/20

394 Reputação

Autoria de 3 guias

Equipe

Cal Poly, Team S1-G1, Zimmerman Spring 2020 Membro de Cal Poly, Team S1-G1, Zimmerman Spring 2020

CPSU-ZIMMERMAN-S20S1G1

Membros da 5

Autoria de 6 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 0

Últimos 7 dias: 1

Últimos 30 dias: 4

Duração total: 140