Ir para o conteúdo principal

Philips Respironics V60 Touchscreen Replacement

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.
  1. Philips Respironics V60 Touchscreen Replacement, Touchscreen: passo 1, imagem 1 %32
    • From the Diagnostics Menu, select Touch Screen Calibration. The Calibrate Touch Screen screen is displayed.

    • Follow the steps shown. Press on the middle of each target with a blunt, narrow object.

    • NOTE: If the Touch Screen Calibration button does not respond, press the Accept button on the navigation ring to begin

    • If the calibration is not successful, have the ventilator serviced.

  2. Philips Respironics V60 Touchscreen Replacement: passo 2, imagem 1 %32
    • Press and release the ON/Shutdown key, opening the Shutdown window in the screen.

    • Select Ventilator Shutdown, which will shut down the ventilator.

    • Improper shutdown may cause a Power has been restored message the next time the ventilator is turned on.

  3. Philips Respironics V60 Touchscreen Replacement: passo 3, imagem 1 %32
    • Remove the screws from the end cap and remove the end cap.

    • Remove the screws from the rear bezel.

    • Separate the front and rear bezel.

  4. Philips Respironics V60 Touchscreen Replacement: passo 4, imagem 1 %32
    • Disconnect these cables from the UI PCBA: • Switch PCBA cable • Nav-ring cable • Power switch overlay cable • Touch screen cable

    • Remove the switch PCBA.

    • Remove the screw that attaches the cable clamp to the LCD tray.

    • Remove the screws and washers that attach the LCD tray to the front bezel, then remove the LCD tray.

    • Remove the touch screen from the front bezel.

Conclusão

To reassemble your device, follow these instructions in reverse order.

Weihan Du

Membro desde: 04/11/20

241 Reputação

Autoria de 2 guias

Equipe

Cal Poly, Team S7-G10, Paton Spring 2020 Membro de Cal Poly, Team S7-G10, Paton Spring 2020

CPSU-PATON-S20S7G10

Membros da 2

Autoria de 9 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 4

Últimos 7 dias: 16

Últimos 30 dias: 42

Duração total: 2,011