  1. Philips Emerson CoughAssist CA Series Disassembly, Philips Emerson CoughAssist CA Series Disassembly: passo 1, imagem 1 %32
    • First, remove the two 16mm pan head screws (1) directly under the front panel of the ventilator. (Page 3).

    • Next, remove the two 16mm flat head screws (3) from the back panel of the ventilator. (Page 3).

    • Lift off the top cover (2) of the device. (Page 3).

  2. Philips Emerson CoughAssist CA Series Disassembly: passo 4, imagem 1 %32 Philips Emerson CoughAssist CA Series Disassembly: passo 4, imagem 2 %32
    • Now remove the "Inhale flow" control knob (4) and "Inhale pressure" knob (4), by loosening the setscrew on each before pulling off. If you have a CM model, carefully remove the Manual Control Lever (3) by prying it off with the screwdriver. (Page 3).

  3. Philips Emerson CoughAssist CA Series Disassembly: passo 5, imagem 1 %32
    • Unscrew the three 12.7mm plastite screws ( 2) holding the panel assembly to the lower housing. A long Phillips screwdriver is very helpful for access to the three screws. (Page 3)

    • The front control panel (5) should now be loose. Carefully slide it forwards to remove. (Page 4).

Conclusão

To reassemble your device, follow these instructions in reverse order.

Zachary Stednitz

Membro desde: 04/07/20

292 Reputação

Autoria de 2 guias

Equipe

Cal Poly, Team S11-G18, Paton Spring 2020 Membro de Cal Poly, Team S11-G18, Paton Spring 2020

CPSU-PATON-S20S11G18

Membros da 3

Autoria de 3 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 0

Últimos 7 dias: 0

Últimos 30 dias: 5

Duração total: 225