Ir para o conteúdo principal

Philips BiPAP S/T Blower Replacement

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.

O que você precisa

Vídeo de Apresentação

  1. Philips BiPAP S/T Blower  Replacement, Blower: passo 1, imagem 1 %32
    • Remove the UI Knob by pulling it straight out.

  2. Philips BiPAP S/T Blower  Replacement: passo 2, imagem 1 %32
    • Remove the top cover by using a Torx T15 to remove the 2 screws on the back of the enclosure.

  3. Philips BiPAP S/T Blower  Replacement: passo 3, imagem 1 %32
    • Remove the Main PCA by removing the wiring harnesses and squeezing the 2 standoffs in the corners of the board.

  4. Philips BiPAP S/T Blower  Replacement: passo 4, imagem 1 %32
    • Remove the Blower Cap by using the T8 Torx screwdriver to remove the 7 screws that secure the Cap to the enclosure.

    • Lift the Blower Assembly out of the blower housing.

    • Remove the blower Outlet Bellows from the blower assembly.

    • Place the New outlet bellows onto the blower assembly.

  5. Philips BiPAP S/T Blower  Replacement: passo 7, imagem 1 %32
    • Align the stands located on the bottom of the blower assembly with the holes in the blower housing.

Conclusão

To reassemble your device, follow these instructions in reverse order.

2 outras pessoas executaram este guia.

Equipe

Cal Poly, Team S7-G3, Paton Spring 2020 Membro de Cal Poly, Team S7-G3, Paton Spring 2020

CPSU-PATON-S20S7G3

Membros da 3

Autoria de 3 guias

Um comentário

this was super helpful. i yanked a working motor/blower from a slightly older system one unit that had a dead humidifier circuit(found it had melted the connector inside) and moved it to my newer bipap unit. works great, dead silent. mine had bad bearings…so loud. great article. thank you for saving me money, time…..SLEEP!

Jon Harmon - Responder

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 2

Últimos 7 dias: 7

Últimos 30 dias: 30

Duração total: 2,762