Newport e360 Oxygen Sensor Replacement

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.

O que você precisa

  1. Newport e360 Oxygen Sensor  Replacement, Oxygen Sensor: passo 1, imagem 1 %32
    • Turn off ventilator before starting this repair.

    • Shut off or disconnect the air and oxygen gas supplies before starting this repair.

  2. Newport e360 Oxygen Sensor  Replacement: passo 2, imagem 1 %32
    • To avoid damaging equipment, always use standard electrostatic discharge (ESD) precautions

    • Put on an ESD wrist strap or ESD gloves

  3. Newport e360 Oxygen Sensor  Replacement: passo 3, imagem 1 %32
    • Remove the 2 screws (phillips size 1) securing DOR2112M to the front of the e360 ventilator.

    • Pull back the cover to gain access to the oxygen sensor

  4. Newport e360 Oxygen Sensor  Replacement: passo 4, imagem 1 %32
    • Remove the Oxygen Sensor cable CBL2109A from the Oxygen Sensor R125P03-002 by unscrewing the connector on the cable

  5. Newport e360 Oxygen Sensor  Replacement: passo 5, imagem 1 %32
    • Remove the Oxygen Sensor by unscrewing the Oxygen Sensor counterclockwise.

    • If the Oxygen Sensor is installed too tightly to be removed, place a Newport reusable circuit on the oxygen sensor to obtain better friction while unscrewing the Oxygen Sensor.

Conclusão

To reassemble your device, follow these instructions in reverse order.

Uma outra pessoa concluiu este guia.

Rylan Loader

Membro desde: 04/09/20

243 Reputação

Autoria de 1 guia

Equipe

Cal Poly, Team S7-G21, Paton Spring 2020 Membro de Cal Poly, Team S7-G21, Paton Spring 2020

CPSU-PATON-S20S7G21

Membros da 3

Autoria de 3 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 0

Últimos 7 dias: 0

Últimos 30 dias: 22

Duração total: 576