Ir para o conteúdo principal

Newport HT70 Pump Assembly Replacement

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.

O que você precisa

  1. Newport HT70 Pump Assembly Replacement, Pump Assembly: passo 1, imagem 1 %32
    • Position the base assembly so that you may access it.

    • Using the #1 Phillips screwdriver, unscrew the four screws (SCR3219P in the photo) and the two screws (SCR3232P) securing the base plate.

  2. Newport HT70 Pump Assembly Replacement: passo 2, imagem 1 %32
    • The base plate will then come apart from the rest of the device with the pump assembly attached to it, as depicted in the following photo.

    • Inspect pump assembly for any damage or looseness.

  3. Newport HT70 Pump Assembly Replacement: passo 3, imagem 1 %32
    • If damage is found, replace the pump assembly by removing each tube individually and replacing as needed.

    • You may not need to replace the entire assembly if parts of it are still intact.

Conclusão

To reassemble your device, follow these instructions in reverse order.