Hamilton-T1 Oxygen Cells Replacement

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.

O que você precisa

  1. Hamilton-T1 Oxygen Cells Replacement, Oxygen Cells: passo 1, imagem 1 %32
    • Flip device around until the back panel is exposed.

  2. Hamilton-T1 Oxygen Cells Replacement: passo 2, imagem 1 %32
    • Remove the screw in the bottom center of the panel, and then take the panel off.

  3. Hamilton-T1 Oxygen Cells Replacement: passo 3, imagem 1 %32
    • Disconnect the oxygen cell by removing the tube.

    • Then twist the cell counter-clockwise and pull to remove the oxygen cell.

    • Install the new oxygen cell and put the back cover on.

    • After the cell is installed. Run the oxygen cell calibration.

    • Oxygen calibration guide can be found in 3-11 of the Hamilton-T1 operations manual.

Conclusão

After the installation and calibration of the oxygen cell. The Hamilton-T1 Ventilator is ready to operate.