  1. Hamilton G5 Safety Valve Block Replacement, Rail: passo 1, imagem 1 %32
    • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.

  2. Hamilton G5 Safety Valve Block Replacement: passo 2, imagem 1 %32
    • Remove the Rail (B).

  3. Hamilton G5 Safety Valve Block Replacement: passo 3, imagem 1 %32
    • The Standoff Brackets (C) remain with the Main Cover.

    • Place aside in a safe location.

  4. Hamilton G5 Safety Valve Block Replacement, Main Cover: passo 5, imagem 1 %32
    • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.

    • Place aside in a safe location.

  5. Hamilton G5 Safety Valve Block Replacement, Safety Valve Block: passo 7, imagem 1 %32
    • Remove Safety Valve Block

    • 1. Disconnect the cable from the Ambient Valve Solenoid

    • 2. Disconnect the Shaped Hose (A)

    • 3. Remove the 2 Hex Allen Screws (B)

    • 4. Remove the Safety Valve Block

    • To remove Ambient Valve Solenoid: Unscrew the Ambient Valve Solenoid (C) form the Safety Valve Block

Conclusão

To reassemble your device, follow these instructions in reverse order.

William

Membro desde: 04/09/20

145 Reputação

Autoria de 2 guias

Equipe

Cal Poly, Team S7-G25, Paton Spring 2020 Membro de Cal Poly, Team S7-G25, Paton Spring 2020

CPSU-PATON-S20S7G25

Membros da 6

Autoria de 11 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 0

Últimos 7 dias: 1

Últimos 30 dias: 3

Duração total: 118