Hamilton G5 Nebulizer Valve Replacement

Equipamento Médico
Nem o iFixit nem este guia de reparo são endossados ou afiliados ao fabricante do dispositivo médico.

O que você precisa

  1. Hamilton G5 Nebulizer Valve Replacement, Rail: passo 1, imagem 1 %32
    • Remove 4 Hex Allen screws (A) from the Rail surrounding the Ventilation Unit.

  2. Hamilton G5 Nebulizer Valve Replacement: passo 2, imagem 1 %32
    • Remove the Rail (B).

  3. Hamilton G5 Nebulizer Valve Replacement: passo 3, imagem 1 %32
    • The Standoff Brackets (C) remain with the Main Cover.

    • Place aside in a safe location.

  4. Hamilton G5 Nebulizer Valve Replacement, Main Cover: passo 5, imagem 1 %32
    • Lift the Main Cover (A) from the Ventilation Unit using the Standoff Brackets (B) attached to the Main Cover.

    • Place aside in a safe location.

  5. Hamilton G5 Nebulizer Valve Replacement, Nebulizer Valve: passo 7, imagem 1 %32
    • After the rail and main cover are taken off, remove the dark blue tubing connected to the nebulizer, pictured here.

  6. Hamilton G5 Nebulizer Valve Replacement: passo 8, imagem 1 %32
    • Remove the two phillips screws labeled A from the Nebulizer valve.

  7. Hamilton G5 Nebulizer Valve Replacement: passo 9, imagem 1 %32
    • Remove the nebulizer valve parts.

    • The Nebulizer Valve parts are:B. Straight fitting

    • C. Front Nozzle fitting

    • D. Elbow fitting

    • E. Mounting bracket

    • F. Hex Allen screw for attaching the bracket to the Safety Valve Block

Conclusão

To reassemble your device, follow these instructions in reverse order.

Cole Korchek

Membro desde: 04/08/20

133 Reputação

Autoria de 1 guia

Equipe

Cal Poly, Team S7-G25, Paton Spring 2020 Membro de Cal Poly, Team S7-G25, Paton Spring 2020

CPSU-PATON-S20S7G25

Membros da 6

Autoria de 11 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 0

Últimos 7 dias: 1

Últimos 30 dias: 5

Duração total: 195