Ir para o conteúdo principal

Autosampler Programming (For Old & New Versions)

O que você precisa

  1. Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: passo 1, imagem 1 %32 Autosampler Programming  (For Old & New Versions), Turn on battery and autosampler: passo 1, imagem 2 %32
    • 1. Switch battery "ON".

    • 2. Autosampler: Push "ON" button.

    • It will say ...STANDBY...

  2. Autosampler Programming  (For Old & New Versions), Select Program Mode: passo 2, imagem 1 %32 Autosampler Programming  (For Old & New Versions), Select Program Mode: passo 2, imagem 2 %32
    • 3. Push "ENTER/PROGRAM" button to see the "[PROGRAM, CONFIGURE] SAMPLER" screen. Push “ENTER/PROGRAM” button to select "PROGRAM".

    • 4. This will bring up the "[TIME, FLOW] PACED SAMPLING" screen. Use arrow to select "FLOW" then push “ENTER/PROGRAM” button.

  3. Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: passo 3, imagem 1 %32 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: passo 3, imagem 2 %32 Autosampler Programming  (For Old & New Versions), Set Number, Type, and Size of Samples: passo 3, imagem 3 %32
    • 5. New screen shows: "SAMPLE EVERY __ PULSES." Press 20. Then push the “ENTER/PROGRAM” button.

    • 6. New screen will read: "MULTIPLEX SAMPLES?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button.

    • 7. New screen will read: "SAMPLE VOLUMES OF "____" ML". Press 1000. Then push the “ENTER/PROGRAM” button.

  4. Autosampler Programming  (For Old & New Versions), Starting Sampling: passo 4, imagem 1 %32 Autosampler Programming  (For Old & New Versions), Starting Sampling: passo 4, imagem 2 %32 Autosampler Programming  (For Old & New Versions), Starting Sampling: passo 4, imagem 3 %32
    • 8. New screen will read: "ENTER START TIME?" Use arrow to select "NO". Then push the “ENTER/PROGRAM” button. New screen: "PROGRAMMING SEQUENCE COMPLETE...". Then screen will read: "PROGRAM HALTED".

    • 9. Push "START SAMPLING" button. Screen will read: "[START, RESUME] SAMPLING PROGRAM?" Push "ENTER/PROGRAM" button to "START". Then push "ENTER/PROGRAM" button again to "START SAMPLING AT BOTTLE 1".

    • 10. Then should see on screen: "START FLOW COUNT AT __:__" (time).

    • It will take some time before it starts sampling.

    • Eventually screen will show: " BOTTLE 1 AFTER 20 PULSES".

Linha de Chegada

Uma outra pessoa concluiu este guia.

Brooke Mason

Membro desde: 05/29/19

2.928 Reputação

Autoria de 17 guias

Equipe

Open-Storm Membro de Open-Storm

Community

Membros da 14

Autoria de 24 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 0

Últimos 7 dias: 0

Últimos 30 dias: 2

Duração total: 164