  1. Autosampler Configuration for Node Control  (For Old & New Versions), Starting the configuration process.: passo 1, imagem 1 %32 Autosampler Configuration for Node Control  (For Old & New Versions), Starting the configuration process.: passo 1, imagem 2 %32 Autosampler Configuration for Node Control  (For Old & New Versions), Starting the configuration process.: passo 1, imagem 3 %32
    • 1. Turn on the autosampler. Screen will read "...STANDBY..."

    • 2. Press "ENTER/PROGRAM" button. Use arrows to highlight "CONFIGURE". Press "ENTER/PROGRAM" button

    • This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the "ENTER/PROGRAM" button.

    • 3. Use arrows to find "EVENT MARK". Press the "ENTER/PROGRAM" button.

  2. Autosampler Configuration for Node Control  (For Old & New Versions), Configuring "event mark".: passo 2, imagem 1 %32 Autosampler Configuration for Node Control  (For Old & New Versions), Configuring "event mark".: passo 2, imagem 2 %32 Autosampler Configuration for Node Control  (For Old & New Versions), Configuring "event mark".: passo 2, imagem 3 %32
    • 4. Use arrow to select "CONTINUOUS SIGNAL". Press "ENTER/PROGRAM" button.

    • 5. Use arrow to select "PUMP CYCLE". Press "ENTER/PROGRAM" button.

    • It is critical this setting is configured in order for the sensor node to operate the autosampler!

    • 6. Use arrows to find "EXIT CONFIGURATION." Press "ENTER/PROGRAM" button. This will bring up the screen "...STANDBY...

Linha de Chegada

Uma outra pessoa concluiu este guia.

Brooke Mason

Membro desde: 05/29/19

2.928 Reputação

Autoria de 17 guias

Equipe

Open-Storm Membro de Open-Storm

Community

Membros da 14

Autoria de 24 guias

0 comentários

Adicione um comentário

Exibir estatísticas:

Últimas 24 horas: 0

Últimos 7 dias: 0

Últimos 30 dias: 0

Duração total: 87