Ir para o conteúdo principal

Vu Surgical Lighting System

42 páginas · 2.58 MB · PDF

SURGICAL LIGHTING SYSTEM

INSTALLATION AND SERVICE MANUAL

Vu Agency Approvals

Medical Electrical Equipment

With respect to electric shock, fire And mechanical hazards only In accordance with IEC60601-1:2006 / CAN/CSA C22.2 No.601.1-M90 with updates 1 & 2

Classifications:

  1. Protection against electrical shock (6.2). Class I permanently connected,
  2. Protection against harmful ingress of water (6.3). None.
  3. Degree of safety in the presence of flammable anesthetics or oxygen (3.11, 3.12, 11.4, 11.5). Not suitable for use in the presence of flammable anesthetics or oxygen.
  4. Mode of operation (6.6). Continuous
  5. Surgical luminaries (IEC60601-2-41)

Electromagnetic compatibility for immunity And emissions in accordance with EN-60601-1-2(2001) Class B and CISPR 22 (1997) Class B

Medical Electrical Equipment Particular requirements for the safety of surgical luminaries and luminaires for diagnosis In accordance with IEC-60601-2-41

Intended use: Professional Medical Lighting for Hospital, Clinic, Minor Surgery, Examination or Diagnosis, within suitable facilities designated for such purposes. This light system is mounted centrally in a room such that access is available all around the operating environment.

User interface: The Vu lighting systems are intended to be used by medical professionals in surgeries and exam rooms. The functional interfaces are up to 360° positioning horizontally (330° for the NVSD3C-E)) and vertically +15° to -65°. The lamp head can be rotated approximately 330°. An Interchangeable Sterilizable Handle assembly provides an ON/OFF button with 5 levels of dimming and Endo mode. An optional Tablet with ON/OFF button, selectable color temperature, 5 levels of dimming, Green Endo mode and Camera control. The NVSD3C-E offers an Interchangeable Camera assembly which replaces the Sterilizable Handle assembly when In-Light Video capabilities are required.

Misuses: For any purpose or use at any facility other than stated above. Caution shall be exercised when positioning the light head and avoid contact or collision with the patent, other medical professionals or other lights/equipment.

User Profile: Any medical professional in surgery or examination rooms.

Vu Installation and Service Manual (1003359 Rev A)

Table of Contents

Table Caption: This table of contents outlines a manual for a surgical lighting system, detailing its specifications, installation procedures, operating instructions, safety guidelines, maintenance, and troubleshooting. It is organized into seven sections covering everything from preliminary information like definitions and symbols to practical aspects of using and maintaining the equipment.

Table of Illustrations

Figure 1: Mounting Dimensions for Ceiling Casting.................................................................12
Figure 2: Ceiling Rod Calculation Solo Mount .........................................................................13
Figure 3: Ceiling Rod Calculation Duo Mount..........................................................................14
Figure 4: Ceiling Rod Calculation Trio Mount ..........................................................................15
Figure 5: Shown, sub-assemblies for a Solo, ceiling mount. ...................................................17
Figure 6: Ceiling Mount............................................................................................................18
Figure 7: Unpacking and installation setup..............................................................................19
Figure 8: Installation setup.......................................................................................................20
Figure 9: Arm and Ceiling Casting Assembly ..........................................................................22
Figure 10: Ceiling & Shaft dowel pin insertion .........................................................................23
Figure 11: Ceiling & Shaft dowel pin locking............................................................................24
Figure 12: Arm Cable termination............................................................................................25
Figure 13: Locking Pin Removal..............................................................................................27
Figure 14: Sterilizable Handle Assembly.................................................................................29
Figure 15: CameraAssembly Figure 16: Inserting Camera Assembly..................................30
Figure 17: Head and Arm Movement.......................................................................................31
Figure 18: Fuse Replacement .................................................................................................34
Figure 19: Arm Adjustment......................................................................................................35
Figure 20: Light Head Adjustment ...........................................................................................36
Figure 21: Arm Friction Adjustment .........................................................................................37
Figure 22: Handle Sterilization.................................................................................................38

Table Caption: This table is a detailed index listing figures and their corresponding page numbers, likely from a technical manual or instruction guide for a medical device, possibly a surgical light or camera system. It provides a visual roadmap to specific components, assembly steps, and maintenance procedures.

Definition of Terms

I.E.C.: International Electrotechnical Commission

ETL: Edison Testing Laboratories

Medical Electrical Equipment: Electrical equipment intended to diagnose or treat a patient under medical supervision. Medical Lighting equipment transfers light energy to an intended area .

Central Illuminance: Illuminance of light head measured at 1 meter from the light emitting area with no